Mechanical installation
18
Load application is implemented via the convex load
button on the top of the force transducer.
The structural component that applies the force to the
convex load application part must be ground and have a
hardness of at least 40 HRC.
HBM provides EDO9 thrust pieces for the C9C, which
are used to keep torques and bending moments away
from the transducer. The thrust pieces naturally have the
correct hardness and surface quality.
Fig. 6.2
Using the EDO9 thrust piece
The substructure must be capable of absorbing the force
to be measured. Remember that the stiffness of the
overall system depends on the stiffness of the force
application part and the substructure. Please also note
that the substructure must ensure that force is always
applied to the transducer vertically, i.e. there must be no
inclination, even under full load.
There are threaded holes underneath the adapter, by
means of which the C9C can also be mounted upside
down or vertically.
A3729-2.2 HBM: public
C9C