Ottobock 1M10 Adjust Gebrauchsanweisung Seite 13

Prothesenfuß
Vorschau ausblenden Andere Handbücher für 1M10 Adjust:
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Alignment reference line of prosthetic foot: between big toe and second toe
   
Alignment reference line of prosthetic socket: along the lateral patella edge
   
Observe the abduction or adduction position.
TF bench alignment
► Observe the information in the prosthetic knee joint instructions for use.
5.1.3 Static Alignment
If necessary, the alignment recommendations (TF modular leg prostheses: 646F219*, TT
modular leg prostheses: 646F336*) may be requested from Ottobock.
Ottobock recommends checking the alignment of the prosthesis using the L.A.S.A.R. Posture
and adapting it as needed.
5.1.4 Dynamic Trial Fitting
Adapt the alignment of the prosthesis in the frontal plane and the sagittal plane (e.g. by mak­
ing angle or slide adjustments) to ensure an optimum gait pattern.
TT fittings:Make sure that physiological knee movement in the sagittal and frontal plane is
achieved when the leg begins to bear weight after the heel strike. Avoid medial movement of
the knee joint. If the knee joint moves in the medial direction in the first half of the stance
phase, medialise the prosthetic foot. If the medial movement occurs in the second half of the
stance phase, reduce the exterior rotation.
5.1.4.1 Optimising the heel characteristics
INFORMATION
Changing the configuration of the function module has no effect on the load line while standing.
The heel characteristics are adjusted with the function module. The following settings are pos­
sible:
Heel strike
Selected configuration (see fig. 2)
Soft
1
Moderate
2 (delivery condition)
Firm
3
► Turn the function module to the desired configuration (see fig. 6).
5.2 Optional: Installing the foam cover
A connecting element (such as a connection plate, connection cap or foam connection cap)
provides a detachable connection between the foam cover and the prosthetic foot.
Required materials: Degreasing cleaner (e.g. 634A58 isopropyl alcohol), 636N9 contact
>
adhesive or 636W17 plastic adhesive
1) Measure the distance from the knee rotation point to the desired end of the foam cover and
add the compression allowance: For PE foam 10 mm, for PUR soft foam 30 mm. For TF
prostheses, add twice the compression allowance.
2) Cut the pre-shaped foam block to length.
3) Pull the foam cover over the prosthesis.
4) Set the connecting element onto the footshell or prosthetic foot. Depending on the version,
the connecting element engages in the edge or sits on the foot adapter.
5) Mount the prosthetic foot on the prosthesis.
6) Mark the outer contour of the connecting element on the distal face of the foam cover.
7) Dismount the prosthetic foot and remove the connecting element.
8) Clean the connecting element using a degreasing cleaner.
Bench alignment process
Position of insert
Insert removed, opening posterior
Insert posterior
Insert anterior
13