Shortening The Cable; Cable Extensions; Parallel Connection (Option); Emc Protection - HBM PW18C3 Serie Montageanleitung

Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

PW18C3... Single point load cells

2.2 Shortening the cable

When the 6−wire configuration is used, the load cell's 6−wire cable can be cut
off without impairing measurement accuracy.

2.3 Cable extensions

You must only use shielded, low-capacitance measurement cables to extend
the cables, making sure that there is a proper connection with minimal contact
resistance. The cable of a six-wire load cell can be extended with a cable of the
same type.

2.4 Parallel connection (option)

Only single point load cells with an aligned output (nominal (rated) sensitivity
and output resistance) are suitable for parallel connection. Many of HBM's
single point load cell types are available with this option.

2.5 EMC protection

Electrical and magnetic fields often induce interference voltages in the
measuring circuit. Therefore:
S Use shielded, low-capacitance measurement cables only
(HBM cables fulfill both conditions)
S Do not route the measurement cables parallel to power lines and control
circuits.
If this is not possible, protect the measurement cable, in rigid steel conduits,
for example
S Avoid stray fields from transformers, motors and contact switches
To ensure the best EMC protection, the load cell, together with the connection
cable and the subsequent electronics, should be placed in a shielded housing.
Exception:
Load cells with shielded, round cables are EMC-tested in accordance with EC
directives and identified by CE certification. But you must make sure that the
shield of the connection cable is connected to the shielding housing of the
electronics.
A2445-1.0 en/de/fr
9