Mechanical installation
5.5
Installing the rotor
Tip
Usually the rotor type plate is no longer visible after
installation. This is why we include with the rotor addi
tional stickers with the important characteristics, which
you can attach to the stator or any other relevant
test‐bench components. You can then refer to them
whenever there is anything you wish to know, such as
the shunt signal. To explicitly assign the data, the identifi
cation number and the size are engraved on the rotor
flange, where they can be seen from outside.
Notice
Make sure during installation that you do not damage the
measuring zone marked in Fig. 5.1 by using it to support
tools, or knocking tools against it when tightening screws,
for example. This can damage the transducer and pro
duce measurement errors, or even destroy the trans
ducer.
T40B
23
A3452-13.0 HBM: public