Orbit OCM160 Betriebsanleitung Seite 65

Inhaltsverzeichnis
OCM160 Precision DC Calibrator
Example:
MEAS:VOLT? Response: 1.000000E+02 V
:MEASure:CURRent[:AMPLitude]?
Syntax:
:MEASure:CURRent[:AMPLitude]?
Description:
This query returns value of output current in Voltage function.
Parameters:
·A
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
MEAS:CURR? Response: 5.000000E-02 A
:MEASure:CONFigure
Syntax:
:MEASure:CONFigure
:MEASure:CONFigure?
Description:
This command sets the function of internal multimeter.
Parameters:
·FREQ
·COUN
Remarks:
The value is stored in each preset
The value is set according to "Startup" preset after power on
The value is set to default on *RST command
Overlapped command
Example:
MEAS:CONF FREQ
MEAS:CONF? Response: FREQ
:MEASure:FREQuency[:AMPLitude]?
Syntax:
:MEASure:FREQuency[:AMPLitude]?
Description:
This query returns value of frequency meter.
Parameters:
·HZ
Remarks:
The value is set to default after power on
The value is set to default on *RST command
Overlapped command
Example:
MEAS:FREQ? Response: 5.000000E+01 HZ
65
Range -5e-2 ... 5e-2
{A}
Amper
{FREQ|COUN} (default FREQ)
Frequency
Counter
Measured value
{HZ}
Hertz