Orbit OCM160 Betriebsanleitung Seite 67

Inhaltsverzeichnis
OCM160 Precision DC Calibrator
Example:
STAT:OPER:ENAB 2
STAT:OPER:ENAB? Response: 2
:STATus:OPERation[:EVENt]?
Syntax:
:STATus:OPERation[:EVENt]?
Description:
This query returns the content of Operational Data Event register. It is a decimal value
which corresponds to the binary-weighted sum of all bits set in the register. Register is
cleared after this query.
Parameters:
Remarks:
Overlapped command
Example:
STAT:OPER? Response: 0
:STATus:OPERation:NTRansition
Syntax:
:STATus:OPERation:NTRansition
:STATus:OPERation:NTRansition?
Description:
This comman allows set Operation Negative Transition Register. It is a decimal value
which corresponds to the binary-weighted sum of all bits set in the register. Setting a bit
in the negative transition filter shall cause a 1 to 0 transition in the corresponding bit of
the associated condition register to cause a 1 to be written in the associated bit of the
corresponding event register.
Parameters:
Remarks:
Overlapped command
Example:
STAT:OPER:NTR 2
STAT:OPER:NTR? Response: 2
:STATus:OPERation:PTRansition
Syntax:
:STATus:OPERation:PTRansition
:STATus:OPERation:PTRansition?
Description:
This comman allows set Operation Positive Transition Register. It is a decimal value
which corresponds to the binary-weighted sum of all bits set in the register. Setting a bit
in the positive transition filter shall cause a 0 to 1 transition in the corresponding bit of
the associated condition register to cause a 1 to be written in the associated bit of the
corresponding event register.
Parameters:
Remarks:
Overlapped command
Example:
STAT:OPER:PTR 1.0
STAT:OPER:PTR? Response: 1.000000E+00
67
Operational Data Event register
Operation Negative Transition Register, Range 0... 32767
Operation Positive Transition Register, Range 0 ... 32767
Orbit Controls AG
Owner's Manual
Inhaltsverzeichnis
loading

Inhaltsverzeichnis