Orbit OCM160 Betriebsanleitung Seite 68

Inhaltsverzeichnis
Orbit Controls AG
:STATus:QUEStionable:CONDition
Syntax:
:STATus:QUEStionable:CONDition?
Description:
This query returns the content of Questionable Condition register. It is a decimal value
which corresponds to the binary-weighted sum of all bits in the register. Register is not
cleared after this query. The response to the query therefore represents an instantaneous
'Snapshot' of the register state, at the time that the query was accepted.
Parameters:
Remarks:
Overlapped command
Example:
STAT:QUES:COND? Response: 2
:STATus:QUEStionable:ENABle
Syntax:
:STATus:QUEStionable:ENABle
:STATus:QUEStionable:ENABle?
Description:
This command enables bits in the Questionable Data Enable register. Selected bits are
summarized at bit 3 (QSS) of the IEEE488.2 Status Byte register.
Parameters:
Remarks:
Overlapped command
Example:
STAT:QUES:ENAB 2
STAT:QUES:ENAB? Response: 2
:STATus:QUEStionable[:EVENt]?
Syntax:
:STATus:QUEStionable[:EVENt]?
Description:
This query returns the content of Questionable Data Event register. It is a decimal value
which corresponds to the binary-weighted sum of all bits set in the register. Register is
cleared after this query.
Parameters:
Remarks:
Overlapped command
Example:
STAT:QUES? Response: 0
:STATus:QUEStionable:NTRansition
Syntax:
:STATus:QUEStionable:NTRansition
:STATus:QUEStionable:NTRansition?
Description:
This comman allows set Questionable Negative Transition Register. It is a decimal value
which corresponds to the binary-weighted sum of all bits set in the register. Setting a bit
in the negative transition filter shall cause a 1 to 0 transition in the corresponding bit of
the associated condition register to cause a 1 to be written in the associated bit of the
corresponding event register.
Owner's Manual
Questionable Condition register
Questionable Data Enable register
Questionable Data Event register
OCM160 Precision DC Calibrator