Since the balance's resolution is high, weighing values may change due to gravity and daily
environmental changes.
4-1
Automatic Calibration (GX-AE/GX-A/GX-AWP Series Only)
The GX-AE/GX-A/GX-AWP series automatically adjusts the balance sensitivity using the built-in weight
according to the temperature change of the operating environment or the set time and interval time. For
details, refer to "7-1 Auto calibration" in the separate instruction manual.
4-2
One-Touch Calibration (GX-AE/GX-A/GX-AWP Series Only)
This function calibrates the balance using the internal mass.
1.
Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the
weighing pan.
2.
Press the CAL
3.
The balance performs calibration using the internal mass. Do not allow vibration or drafts to affect
the balance.
4.
After the calibration, if the "GLP output (
outputs a sensitivity adjustment execution record.
5.
The balance will automatically return to weighing mode after calibration.

4-3 Calibration Using an External Weight

This function calibrates the balance using an external weight.
1.
Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the
weighing pan.
2.
Press and hold the CAL key for 2 seconds until
3.
Make sure that nothing is on the weighing pan and press the
Do not apply vibration, etc.
4.
Place the external weight on the weighing pan and press the
etc.
5.
Remove the external weight from the weighing pan.
6.
After calibration, if GLP output is set, "sensitivity adjustment execution record" is output or stored in
data memory.
7.
The display automatically returns to weighing display.
8.
Place the external weight again and check that the set value is ± 2 digits.
If it is out of range, pay attention to the surrounding environment and start from "1".
4. Calibration
key. The balance displays
.
)" parameter of the function table is set, the balance
is displayed, then release the key.
PRINT
PRINT
10
key to weigh the zero point.
key. Do not apply vibration